Simultaneous determination of phenylureas and triazines in water, by capillary gas chromatography with ion trap detection and large volume injection
C. Charrêteur1, R. Colin1, D. Morin2 and J.J. Péron21 Laboratoire d'Hygiène Publique et Industrielle, Université de Bretagne Sud, Centre de Génie Industriel, Guidel-plages, 56520 Guidel, France
2 Laboratoire de Chimie Analytique Appliquée, Université de Bretagne Sud, Centre de Génie Industriel, Guidel-plages, 56520 Guidel, France
Abstract
A GC method is described for the simultaneous determination of triazines and phenylureas in aqueous samples by large volume injection with a Programmable Temperature Vaporisator injector
(PTV) and ion trap detection (ITD). The analytical column (DB 5MS, 30 m, 0.25 mm, 0.25 mm) is connected to a pre-column (DB 5MS, 5 m, 0.25 mm, 0.25 mm) via a cross piece valve (MCS) to
prevent residual solvant vapour from polluting the analytical column. The 8 pesticides (4 triazines and 4 ureas) are determined by this method, the linearity is improved using 2 Fisher test. The
detection and the quantification limits are also determined for each compounds and are less than 0.10 mg/L which is the limit fixed by the ECC reglementation on drinking water.